11-feb-13 02:50:00 Measurements ended --- . Number of lamps (L) = 7 . Filter running time = 32.75 hours . Total airflow this filter = 9.59 cubic meters . Mean BC concentration of all lamps = 147 ng/m3 . Report for lamp 1 (370 nm): . Ending voltages: sen(beam) = .506 sen(zero) = 0.021 . ref(beam) = 2.163 ref(zero) = 0.021 . Optical attenuation of filter deposit = 99 . Total aerosol black carbon on filter = 1254 ng Mean aerosol black carbon concentration = 131 ng/m3 --- The reference beam showed lamp intensity fluctuations of 34 ppm. --- Report for lamp 2 (470 nm): . Ending voltages: sen(beam) = .932 sen(zero) = 0.021 . ref(beam) = 2.377 ref(zero) = 0.021 . Optical attenuation of filter deposit = 85 . Total aerosol black carbon on filter = 1368 ng Mean aerosol black carbon concentration = 142 ng/m3 --- The reference beam showed lamp intensity fluctuations of 25 ppm. --- Report for lamp 3 (520 nm): . Ending voltages: sen(beam) = .667 sen(zero) = 0.021 . ref(beam) = 2.420 ref(zero) = 0.021 . Optical attenuation of filter deposit = 76 . Total aerosol black carbon on filter = 1366 ng Mean aerosol black carbon concentration = 142 ng/m3 --- The reference beam showed lamp intensity fluctuations of 26 ppm. --- Report for lamp 4 (590 nm): . Ending voltages: sen(beam) = .981 sen(zero) = 0.021 . ref(beam) = 1.240 ref(zero) = 0.021 . Optical attenuation of filter deposit = 70 . Total aerosol black carbon on filter = 1412 ng Mean aerosol black carbon concentration = 147 ng/m3 --- The reference beam showed lamp intensity fluctuations of 88 ppm. --- Report for lamp 5 (660 nm): . Ending voltages: sen(beam) = 1.254 sen(zero) = 0.021 . ref(beam) = 2.275 ref(zero) = 0.021 . Optical attenuation of filter deposit = 65 . Total aerosol black carbon on filter = 1479 ng Mean aerosol black carbon concentration = 154 ng/m3 --- The reference beam showed lamp intensity fluctuations of 51 ppm. --- Report for lamp 6 (880 nm): . Ending voltages: sen(beam) = 1.432 sen(zero) = 0.021 . ref(beam) = 3.505 ref(zero) = 0.021 . Optical attenuation of filter deposit = 50 . Total aerosol black carbon on filter = 1521 ng Mean aerosol black carbon concentration = 158 ng/m3 --- The reference beam showed lamp intensity fluctuations of 51 ppm. --- Report for lamp 7 (950 nm): . Ending voltages: sen(beam) = 1.165 sen(zero) = 0.021 . ref(beam) = 4.516 ref(zero) = 0.021 . Optical attenuation of filter deposit = 46 . Total aerosol black carbon on filter = 1512 ng Mean aerosol black carbon concentration = 157 ng/m3 --- The reference beam showed lamp intensity fluctuations of 49 ppm. --- Disk space = 2.69993E+07 bytes free: One entry at every 5 minutes. There is space for 5740 hours ( 239 days) more data --- 11-feb-13 02:50:00 Tape feeder mechanism advancing for 2 spot(s). --- ============================================================= --- MAGEE SCIEN. AETHALOMETER --- Aethalometer No. 463 --- 11-feb-13 03:04:54 Measurements started Starting voltages: . sen(zero) = 0.021 ref(zero) = 0.021 . Lamp 1 (370 nm): sen(beam) = 1.309 ref(beam) = 2.164 . Lamp 2 (470 nm): sen(beam) = 2.137 ref(beam) = 2.390 . Lamp 3 (520 nm): sen(beam) = 1.402 ref(beam) = 2.447 . Lamp 4 (590 nm): sen(beam) = 1.920 ref(beam) = 1.240 . Lamp 5 (660 nm): sen(beam) = 2.377 ref(beam) = 2.285 . Lamp 6 (880 nm): sen(beam) = 2.342 ref(beam) = 3.519 . Lamp 7 (950 nm): sen(beam) = 1.819 ref(beam) = 4.516 ---