OutageID |
SiteID |
Failure Type |
Start Date |
End Date |
Description |
Resolution Description |
871 |
ctd |
Hardware - Instrument |
2023-08-15 05:40:00 |
2023-09-01 01:55:35 |
Radar PC unable to communicate with the disdrometer |
Connection reestablished |
868 |
ctd |
Hardware - Instrument |
2023-08-01 18:58:00 |
2023-09-21 20:04:00 |
Unable to connect to datalogger from the radar PC. Possible data cable or connector issue. |
Data connection restored |
858 |
ctd |
Power Failure |
2023-06-04 22:00:00 |
2023-06-12 20:00:00 |
Site power has been disabled by the utility to troubleshoot power issues around Courtland. |
Power issues resolved. |
829 |
ctd |
Power Failure |
2022-11-26 09:00:00 |
2022-12-10 03:45:00 |
Unable to connect to cellular router; possible power or router failure |
Power restored |
814 |
ctd |
Hardware - Instrument |
2022-05-11 23:00:00 |
2022-06-27 23:49:54 |
Disdrometer RS-485/RS-232 converter malfunction; lab repair required |
Disdrometer replaced |
811 |
ctd |
Hardware - Instrument |
2022-05-02 20:45:00 |
2022-05-10 20:45:00 |
Recurring 915-MHz transmitter issue |
PIRAQ card swapped out |
808 |
ctd |
Hardware - PC |
2022-04-15 21:15:00 |
2022-04-20 17:30:00 |
S-band computer hard disk failure |
Hard disk replaced |
794 |
ctd |
Hardware - Instrument |
2021-12-07 02:30:00 |
2022-04-26 23:00:00 |
915-MHz radar transmitter issue |
Transmitter repaired and installed |
747 |
ctd |
Hardware - Instrument |
2020-12-10 04:00:00 |
2021-03-16 18:30:00 |
Radar DSP controller card malfunction |
Radar electronics/PC replaced |
683 |
ctd |
Hardware - Instrument |
2020-01-19 23:00:00 |
2020-02-28 23:00:00 |
Radar electronics failure; engineering site visit scheduled for the week of Jan 27, 2020 |
Radar PC and RASS controller card replaced |
681 |
ctd |
Hardware - Instrument |
2020-01-15 02:08:00 |
2020-02-03 18:00:00 |
Site PC can't communicate with datalogger over wireless modems; datalogger voltage (10.106) was at a critically low operating threshold before outage |
Power now supplied by AC; RF modems replaced with data cable; datalogger has intermittent communication issues; data acquisition software was modified with enhanced fault-tolerance methods to handle the datalogger communication faults |