OutageID |
SiteID |
Failure Type |
Start Date |
End Date |
Description |
Resolution Description |
864 |
bby |
Hardware - Instrument |
2023-02-14 00:00:00 |
2023-08-10 00:16:00 |
Anemometer sensor malfunction |
Sensor repaired |
869 |
bby |
Hardware - Instrument |
2022-05-26 00:00:00 |
2023-09-13 02:28:00 |
Relative humidity sensor malfunction |
Sensor replaced |
752 |
bby |
Hardware - Instrument |
2021-01-05 00:00:00 |
2021-01-06 18:00:00 |
Tipping bucket grossly under-reporting precipitation amounts; begin time of malfunction unknown |
Funnel assembly blown off tipping bucket by wind; funnel back in place |
606 |
bby |
Hardware - Instrument |
2018-09-11 20:00:00 |
2018-09-13 23:30:00 |
On-site radar hardware maintenance. |
Switches replaced. |
514 |
bby |
Hardware - Instrument |
2017-01-08 04:00:00 |
2017-01-19 18:00:00 |
A/C fan motor needs to be replaced. Replacement for parts was placed on 1/9/2017. |
AC issue still pending resolution. System is running and collecting data, in the mean time. |
502 |
bby |
Hardware - Instrument |
2016-09-12 00:00:00 |
2016-11-10 22:00:00 |
Net Radiation Dome broke. Need to replace next visit. |
Net Rad Fixed. |
463 |
bby |
Hardware - Instrument |
2015-09-09 22:00:00 |
2015-09-11 00:00:00 |
AC unit down. Powered off radar to prevent damage from overheating. |
AC repaired. |
406 |
bby |
Hardware - Instrument |
2014-10-03 19:00:00 |
2014-10-15 00:00:00 |
Radar exceeded transmitter temperatures thresholds. Radar shut down to prevent any damage. |
Transmitter temperature sensors may need replacing. Issue has been resolved for the time being. |
364 |
bby |
Hardware - Instrument |
2014-02-11 23:00:00 |
2014-02-24 16:00:00 |
AC unit down. Powered off radar to prevent damage from overheating. Updated 2/21/14. New motor to be replaced Monday for AC unit. |
AC motor replaced. |
344 |
bby |
Hardware - PC |
2013-09-07 00:00:00 |
2013-11-07 00:00:00 |
Unable to communicate with datalogger. COM port okay on PC. |
Conduit and cable was damaged. Repairs made. |
311 |
bby |
Hardware - Instrument |
2013-05-24 00:00:00 |
2013-07-17 18:00:00 |
Rain gauge switched may have malfunctioned causing false precipitation to be reported. |
Rain gauge fixed. |
306 |
bby |
Hardware - Instrument |
2013-04-15 00:00:00 |
2013-05-08 16:00:00 |
Rain gauge switched may have malfunctioned causing false precipitation to be reported. |
Fixed. |
272 |
bby |
Communication - Cellular |
2012-11-13 22:00:00 |
2013-03-22 00:00:00 |
Site removed until new 449 MHz replacement. |
New 449 MHz site installed. |
260 |
bby |
Hardware - Instrument |
2012-08-15 02:00:00 |
2012-10-02 02:00:00 |
Datalogger unreachable from computer on site. Possible data cable or datalogger problem. |
Datalogger connection restored after PC reboot. |
247 |
bby |
Hardware - Instrument |
2012-07-05 21:00:00 |
2012-10-11 00:00:00 |
Profiler getting no return signal. Investigating cause. Update 7/10/12. bad fuse was replaced but profiler still getting no return. Could be problem with transmitter. Profiler will be shut down until repairs are made. |
Profiler repaired on site by engineer. |
91 |
bby |
Hardware - Instrument |
2011-05-04 17:00:00 |
2011-09-08 01:00:00 |
Net radiometer removed from site. Values being reported must be discarded. Replacement net radiometer will be installed during next site visit. |
Net radiometer replaced. |
23 |
bby |
Hardware - Instrument |
2011-01-06 18:36:42 |
2011-03-06 18:36:42 |
Backup APC critical failure. Radar system has by-passed UPS. Recommend new UPS. |
New UPS installed. |
15 |
bby |
Hardware - Instrument |
2010-12-16 19:00:00 |
2011-01-07 00:00:00 |
Reported problems with RASS. |
Power to RASS was unplugged. Plugged back in to restore power. |
12 |
bby |
Communication - Phone |
2010-12-16 18:00:00 |
2011-01-07 00:00:00 |
Reported problems with telephone line. Tested too much AC Voltage. |
Installed a cell phone modem to by-pass phone line. Phone modem is still connected to the PC and will serve as a back-up. |
10 |
bby |
Communication - Phone |
2010-12-07 22:00:00 |
2010-12-10 00:00:00 |
Phone line was down at D-mark. Phone company working fixing the problem. |
Phone line restored at D-mark. |