| OutageID |
SiteID |
Failure Type |
Start Date |
End Date |
Description |
Resolution Description |
| 959 |
str |
Hardware - Instrument |
2025-07-08 11:00:16 |
|
Disdrometer does not respond to communication. Outage start corresponds to the data drive filling up with radar data, but fixing that has not fixed the issue. Unclear if that is the actual cause. There is also an issue with the PIRAQ on the radar. |
|
| 916 |
str |
Hardware - PC |
2024-06-09 22:58:00 |
2024-12-03 20:00:00 |
Radar computer not booting |
Replaced computer |
| 934 |
str |
Hardware - Instrument |
2024-06-09 22:58:00 |
2025-07-17 10:57:16 |
Unable to connect to datalogger. Site visit showed that the cable had been cut. Need new cable. |
Jackson replaced the met cable |
| 884 |
str |
Hardware - Instrument |
2023-10-18 02:15:00 |
2024-01-26 00:00:00 |
Receiver modulator box malfunction |
Cold solder joint repair |
| 907 |
str |
Hardware - Instrument |
2023-10-18 02:15:00 |
2024-03-13 21:15:00 |
Coupler disabled |
Coupler enabled |
| 857 |
str |
Hardware - Instrument |
2023-05-13 02:00:00 |
2023-07-12 17:54:00 |
Low battery voltage preventing overnight datalogger operations |
Battery replaced |
| 855 |
str |
Hardware - Instrument |
2023-04-23 22:16:00 |
2023-07-12 17:54:00 |
Temperature/relative-humidity sensor malfunction |
Sensor replaced |
| 856 |
str |
Hardware - Instrument |
2023-04-14 07:34:00 |
2023-07-12 17:54:00 |
Pressure sensor malfunction |
Sensor replaced |
| 612 |
str |
Power Failure |
2018-11-21 16:00:00 |
2019-04-24 21:40:00 |
Power issue confirmed. |
Power restored. |
| 590 |
str |
Hardware - Instrument |
2018-04-30 21:12:00 |
2018-05-10 21:00:00 |
Can't connect to datalogger. Data cable from PC to datalogger was damaged during groundskeeping work. |
Cable replaced. |
| 558 |
str |
Communication - Cellular |
2017-10-09 13:00:00 |
2017-10-11 03:45:00 |
Router unreachable |
Cellular service restored. Wildfire in the vicinity. |
| 741 |
str |
Hardware - Instrument |
2016-11-29 23:00:00 |
2022-06-14 02:40:00 |
100-cm soil reflectometer data is suspect |
sensor verified to be working correctly; suspect data may be related to abnormal soil and/or water behaviors |
| 503 |
str |
Communication - Cellular |
2016-10-06 12:00:00 |
2016-10-10 00:00:00 |
Lost connectivity to radar PC via cell modem. |
Comms restored. |
| 486 |
str |
Communication - Cellular |
2016-05-28 14:00:00 |
2016-06-01 22:45:00 |
Unable to reach router. Possible power outage or router issue. |
UPS reset and power restored. |
| 414 |
str |
Communication - Cellular |
2014-10-24 00:00:00 |
2014-11-05 21:00:00 |
Router not responding. Could be a router and/or power issue. |
Contact rebooted system. |
| 369 |
str |
Hardware - Instrument |
2014-03-07 00:00:00 |
2014-04-07 19:00:00 |
Datalogger is dead. Needs to be replaced. |
Datalogger replaced. |
| 363 |
str |
Hardware - Instrument |
2014-02-07 00:00:00 |
2014-02-27 00:00:00 |
Tipping bucket gauge most likely clogged as seen from linear rain trace in plot. |
clogged fixed |
| 188 |
str |
Hardware - Instrument |
2012-02-08 00:00:00 |
2012-02-15 00:00:00 |
Faulty Temperature/RH sensor. Reporting values too low. Sending contact replacement sensor. |
Replaced Temperature/RH sensor. |